Functions and Measurements
Making Distortion Measurements
Notice the
Marker
D
amplitude readout. This is the difference
of the distortion product amplitude between 0 dB and 10 dB
input attenuation settings. If the
Marker
Δ
absolute amplitude
is approximately
1 dB for an input attenuator change, the
analyzer is generating, at least in part, the distortion.
The
Marker
Δ
amplitude readout comes from two sources:
1)
Increased input attenuation causes poorer signal- to- noise
ratio. This causes the
Marker
Δ
to be positive.
2)
The reduced contribution of the analyzer circuits to the
harmonic measurement causes the
Marker
Δ
to be negative.
Large
Marker
Δ
readout indicates significant measurement
errors. Set the input attenuator to minimize the absolute
value of
Marker
Δ
.
≥