Inspection, Installation, and Verification
1
Keysight M9260A Startup Guide
35
Conduct operational verification (optional)
The verification procedure in this section helps to ensure the M9260A module is
operational. No external equipment is required except for two BNC to BNC cables
to enable loopback configuration from the audio generator outputs to the audio
analyzer inputs.
M9260A operational verification procedure
The M9260A verification procedure involves measuring the Vac, Frequency,
THD+N, and THD parameters. The average value from the measured data samples
will then be calculated. The measurement results (average) are then compared to
the known values output from the audio generator.
1
Launch the M9260A Soft Front Panel (SFP).
2
Connect the Generator CH1 output to the Analyzer CH1 input, and the
Generator CH2 output to the Analyzer CH2 input using BNC cables as shown
in
NOTE
– The SFP can be used as a learning tool to help you experience the module's
capability and behavior through an easy-to-use GUI.
– Refer to the Keysight M9260A Soft Front Panel Help for details on how to use
the SFP software.