Keysight E8363B/C and E8364B/C Installation Note 15
General Information
Installation Procedure for the Upgrade Kit
Step 5.Install the Attenuators and Bias Tees
Refer to
for this portion of the procedure.
To install the attenuators and bias tees, the brackets holding the detectors
must be removed.
1.
With a 5/16-inch torque wrench, disconnect, remove, and keep for
reinstallation later, the two semirigid cables (item
①
) from each detector.
For analyzers without Option 014 installed, one of these cables has already
been removed.
2.
With a T-10 TORX driver, remove the three mounting screws (item
✍
) from
each bracket holding the detector.
3.
Remove the brackets from the analyzer with the detectors attached.
Figure 8
Attenuator and Bias Tee Installation, Bracket Removal
Refer to
for this portion of the procedure.