Keysight E8363B/C and E8364B/C Installation Note 7
General Information
Installation Procedure for the Upgrade Kit
Step 1.Remove the Outer Cover
Refer to
for this procedure.
1.
Disconnect the power cord (if it has not already been disconnected).
2.
With a T-20 TORX driver, remove the strap handles (item
①
) by loosening
the screws (item
✍
) on both ends until the handle is free of the analyzer.
3.
With a T-20 TORX driver, remove the four rear panel feet (item
③
) by
removing the center screws (item
④
).
4.
Slide the four bottom feet (item
⑤
) off the cover.
5.
Slide the cover off of the frame.
This procedure is best performed with the analyzer resting on its front
handles in the vertical position.
Do not place the analyzer on its front panel
without the handles
. This will damage the front panel assemblies.