42
8.2 Configuration settings
Note - when changing Protocols, the AFA1000 should be power cycled once re-configured to ensure
the changed take effect.
The TEL protocol has fixed parameters so cannot be adjusted in the field, to select the TEL protocol
using the pushbutton menus: -
1.
Press Enter from the “Requires set up” screen or if the monitor is in the Run Screen Press and
Hold the Enter button for 5 seconds until the Main Menu is displayed.
2.
Using the + / - buttons select SET UP, and then select CONFIGURE, then enter the password (the
factory default Password is 0-0-0-0), press Enter to continue.
3.
Using the + / - buttons select PROTOCOL and press Enter, then using the +/- buttons select TEL
and then press Enter to continue, then select DONE to return to the CONFIG MENU, then select
DONE and RUN.
To Setup the AFA1000 with MODBUS protocol using the pushbutton menus: -
1.
Press Enter from the “Requires set up” screen or if the monitor is in the Run Screen Press and
Hold the Enter button for 5 seconds until the Main Menu is displayed.
2.
Using the + / - buttons select SET UP, and then select CONFIGURE, then enter the password (the
factory default Password is 0-0-0-0), press Enter to continue.
3.
Using the + / - buttons select PROTOCOL, and press Enter, then using the +/- buttons select
MODBUS and then press Enter to return to the CONFIG MENU.
4.
Using the + / - buttons select MODBUS SETTINGS and then press Enter, a sub menu will appear
with the following parameters: -
5.
Using the + / - buttons select the SLAVE ID parameter and then press Enter, then using the +/-
buttons select the required ID and press Enter to continue.
6.
Using the + / - buttons select the BAUD RATE parameter and then press Enter, then using the +/-
buttons select the required rate and press Enter to continue.
7.
Using the + / - buttons select the PARITY parameter and then press Enter, then using the +/-
buttons select the required rate and press Enter to continue then select DONE to return to the
CONFIG MENU, then select DONE and RUN.
Parameter
Description
Range
Default
A Slave ID
Set the slave ID for
the unit
1-255
1
B Baud Rate
Set the network
Baud Rate
1200/2400/4800/9600/1440/19200/38400/57600
9600
C Parity
Set the required
parity
None/Even/Odd
None
Summary of Contents for Air Alert 600
Page 2: ...2 Issue 02 June 2017...
Page 5: ...5 1 SAFETY...
Page 10: ...10 2 Overview of the Airflow Monitor...
Page 20: ...20 3 Functions Operation...
Page 25: ...25 4 Components...
Page 28: ...28 5 Installation...
Page 33: ...33 6 Configuration Calibration...
Page 37: ...37 7 Auxiliary Features Connections...
Page 40: ...40 8 RS485 Coms Output...
Page 47: ...47 9 Warranty...