22
3.2 Input Functions.
The airflow monitor has 3 programmable Inputs that can be set to analogue (0-5VDC), digital open or
digital closed operation.
Analogue input functions
Temperature
•
Any Input set to Analogue - Temperature
•
Temperature is displayed alongside airflow velocity in ͦF or ͦC
•
Low Temp toggles on / off with display if temperature < Low Temp Alarm point
•
Screen Backlight Red
•
Audible alarm sounds -- can be muted via Enter pushbutton
•
Low Temp relay operates (if configured)
•
High Temp toggles on / off with display if temperature > High Temp Alarm point
•
Screen Backlight Red
•
Audible alarm sounds -- can be muted via Enter pushbutton
•
High Temp relay operates (if configured)
Digital input functions
Alarm disable
•
When input configured as Alarm disable is activated
•
Alarm disabled is displayed
•
Screen Backlight Red
•
Audible alarm muted
•
Mute Icon shown on display
Night set-back
•
When input configured as Night set-back is activated
•
Night Set-back Icon displayed
•
Audible alarm muted
•
Mute Icon shown on display
•
Low Air alarm muted if set to Maintain low air
•
Reduced Low Air & Warning alarms active if set to Reduce low air
•
Screen Backlight Red and Backlight Dims