3.1 Airflow Functions.
The airflow monitor velocity display can be set up using the pushbutton menus to display airflow in
units of m/sec or fpm and can also be set to show plain text "Air Safe" & "Air Fail" only.
The airflow monitor has 3 programmable airflow alarms.
Safe airflow
•
Airflow reading above warning level (e.g. > 0.90 fpm)
•
Screen Backlight Green
Warning airflow
•
Airflow reads between warning level and air fail level (e.g. > 80 fpm and < 90 fpm)
•
Screen Backlight Amber
Low airflow
•
Airflow reads below alarm level for longer than the warning to low air delay time
•
Air Fail toggles on / off with display
•
Screen Backlight Red
•
Audible alarm sounds -- can be muted via Enter pushbutton
•
Low air relay operates (if configured)
Reset: -- when airflow rises 3 fpm above Low air level for longer than the low air to
warning air delay time the Low air alarm resets automatically
High airflow
If configured:
•
Airflow reading above high level (e.g. > 300 fpm)
•
High Air toggles on / off with display
•
Screen Backlight Red
•
High air relay operates (if configured)
Audible Alarm Mute
When the audible alarm is muted via the Enter button - an Icon (horn with forward slash) is shown on
the display
.
The audible alarm can be permanently disabled in the pushbutton menu.
Summary of Contents for Air Alert 600
Page 2: ...2 Issue 02 June 2017...
Page 5: ...5 1 SAFETY...
Page 10: ...10 2 Overview of the Airflow Monitor...
Page 20: ...20 3 Functions Operation...
Page 25: ...25 4 Components...
Page 28: ...28 5 Installation...
Page 33: ...33 6 Configuration Calibration...
Page 37: ...37 7 Auxiliary Features Connections...
Page 40: ...40 8 RS485 Coms Output...
Page 47: ...47 9 Warranty...