29
5.1 Installation of standard components
The following section outlines the installation of the various components of the Airflow monitor.
As the size and format of individual Fume Hoods varies considerably, specific instructions are not
possible, though the principles outlined below should remain valid in all cases.
The airflow monitor can be mounted on either side of the Fume Hood; consideration should be made
for the cable lengths when fitting the Airflow sensor.
1. Fit the AA600 monitor to the Fume Hood using the cut-out details provided with the unit --- see page
29. Ensure that the monitor fits into the cut out before securing with the fixing screws.
Separate mounting boxes and adaptor plates are available for retro-fitting to older Fume Hoods.
2. Fit the airflow sensor to the Fume Hood using the cut out and installation details provided --- see
page 30 & 31. When possible ensure that the Airflow sensor is mounted on the same side of the Fume
Hood as the AA600 so that the standard 2M/78” sensor cable will reach the monitor.
4. Connect the ‘telephone style’ airflow sensor plug-in cable to the back of the AA600 unit --- see
typical connection diagram on page 25
Longer Sensor Cables are available on request.