35
6.3 Calibration
1.
Press Enter from the “Requires set up” screen or if the monitor is in the Run screen Press and
Hold the Enter button for 5 seconds until the Main Menu is displayed.
2.
Using the + / - buttons select SET UP, then select CALIBRATION, then enter the password (the
factory default Password is 0-0-0-0), press Enter to continue.
3.
With the sash open to the normal operating height again measure the face velocity using a
calibrated instrument. Using the + / -buttons enter the measured face velocity then press Enter,
the monitor will then sample the airflow for 5 seconds.
4.
If the airflow sample is unstable the monitor will display “Deviations too High”, follow the
instructions to repeat the sample or quit the calibration.
5.
If the sample is accepted, lower the sash by half and measure the face velocity using a calibrated
instrument. Using the + / - buttons enter the measured face velocity and press Enter, the monitor
will then sample the airflow for 5 seconds.
6.
If the calibration is successful the monitor will return the Main Menu, select RUN to go to normal
operating mode and check the velocity reading is accurate and stable
.
7.
If the airflow sample is unstable the monitor will display “Deviations too High”, follow the
instructions to repeat the sample or quit the calibration.
8.
The monitor will display “Increase higher airflow sample” if the second sample value is too close to
the first value entered, close the sash a little and repeat the higher sample. The minimum
difference between the samples that the monitor will accept is 0.3m/sec (60fpm).
9.
The monitor will display “Sensor diff too low” if the monitor doesn’t detect any difference in the
sensor output between the 2 airflow samples, check that the sensor hose is connected and repeat
the calibration.
10.
If the calibration is stable and accurate select RUN from the Main Menu to go to normal operating
mode.
See Calibration Notes for hints on successful calibration.
Summary of Contents for Air Alert 600
Page 2: ...2 Issue 02 June 2017...
Page 5: ...5 1 SAFETY...
Page 10: ...10 2 Overview of the Airflow Monitor...
Page 20: ...20 3 Functions Operation...
Page 25: ...25 4 Components...
Page 28: ...28 5 Installation...
Page 33: ...33 6 Configuration Calibration...
Page 37: ...37 7 Auxiliary Features Connections...
Page 40: ...40 8 RS485 Coms Output...
Page 47: ...47 9 Warranty...