Instructions for use elements™ 8:1
5 Operation | 5.3 Inserting the dental bur or diamond grinder
WARNING
Use of unauthorised dental burs.
Risk of injury.
▶
Comply with the instructions for use and the intended use of the dental bur.
▶
Only use dental bur that do not deviate from the specified data.
WARNING
Hazard from rotating dental bur.
Lacerations.
▶
Do not touch the rotating dental bur!
▶
Remove the dental burs from the medical device after treatment to avoid
injury and infection during storage.
CAUTION
Dental bur with worn or damaged shafts.
Risk of injury, dental bur may fall out during treatment.
▶
Never use a dental bur with damaged or worn shafts.
CAUTION
Hazard from defective chuck system.
The dental bur can fall out and cause injury.
▶
Pull on the dental bur to check if the clamping system works properly and if
the dental bur is firmly clamped. Wear gloves or a thimble when you check,
insert, or remove the bits to prevent injury and infection.
NOTICE
Dental bur shaft slips inside the chuck due to excessive speed of the
dental bur or abrupt engagement of the dental bur.
Material damage to dental bur shaft and chuck system, reduction of the service
life of dental bur and chuck system.
▶
Do not operate the dental bur at a higher speed than recommended by the
manufacturer.
NOTICE
Dental bur with worn or damaged shafts.
Material damage to the chuck system, dental bur is difficult or impossible to re-
move from the chuck system.
▶
Never use a dental bur with damaged or worn shafts.
▶
Insert the dental bur into the segment of the head drive by twisting the
dental bur slightly, and push it to the end stop.
▶
Check if the dental bur is seated securely by pulling on it.