7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
7.4 Care products and systems - Servicing
WARNING
Sharp tool in the medical device.
Injury hazard from sharp and/or pointed tool.
▶
Remove the tool.
CAUTION
Improper service and care.
Risk of injury.
▶
Service regularly with suitable agents.
Note
KaVo guarantees the proper function of KaVo products only if the care
products listed as accessories are used, since these were tested for proper
use on our products.
7.4.1 Servicing with KaVo Spray
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Remove the tool from the medical device.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.
▶
Press the spray key for 1 to 2 seconds.
Servicing the chucking system
KaVo recommends servicing the chucking system once weekly.
▶
Remove the tool from the medical device.
▶
Position the tip of the spray nipple in the opening, and apply the spray.
▶
Press the spray key for 1 to 2 seconds.
7.4.2 Servicing with KaVo QUATTROcare PLUS
Servicing and cleaning device with expansion pressure for the interior cleaning
of inorganic residues and optimum care.
(no validated cleaning of the interior in accordance with German RKI require-
ments)
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Remove the tool from the medical device.
25 / 32