Instructions for use elements™ 8:1
7 Reprocessing steps in accordance with ISO 17664 | 7.4 Care products and systems - Servicing
26 / 32
▶
Service the product in the QUATTROcare PLUS.
See also:
2
Instructions for use KaVo QUATTROcare PLUS
Servicing the chuck
KaVo recommends servicing the chuck system once a week using the chuck
servicing program integrated in the device.
Note
Handpieces must be taken off the service couplings before the chuck service
can be started and performed.
▶
Close the front door and press the chuck service button for at least three
seconds until the spray canister control LED flashes three times consecut-
ively.
ð
The device is in chuck service mode.
▶
Remove the service coupling of the chuck from the side hatch of the QUAT-
TROcare PLUS and attach it to coupling service point four, on the far right.
A MULTIflex adaptor must be mounted there.
▶
Press the handpiece together with the guide bush of the chuck to be ser-
viced against the tip of the service coupling.
▶
Press the button marked with the chuck service symbol.
Note
Close the chuck service mode.
Option 1: Place the dental handpieces in the QUATTROcare PLUS 2124 A,
close the front door and start theservice procedure.
Option 2: After three minutes with no service procedure running, the device
automatically switches back to normal service mode.
See also:
2
Servicing with KaVo QUATTROcare PLUS
7.4.3 Care with KaVo SPRAYrotor
Note
KaVo SPRAYrotor is no longer included in the current delivery pro-
gramme.
Follow-up product:
▶
QUATTROcare PLUS 2124 A
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Cover the medical device with the Cleanpac bag, and place it on the corres-
ponding servicing adapter on the KaVo SPRAYrotor.
▶
Service the product.