Instructions for use elements™ 8:1
7 Reprocessing steps in accordance with ISO 17664 | 7.5 Packaging
See also:
2
Instructions for use KaVo SPRAYrotor
7.4.4 Servicing with KaVo QUATTROcare
Note
QUATTROcare 2104 / 2104 A is no longer included in the current de-
livery programme.
Follow-up product:
▶
QUATTROcare PLUS 2124 A
Servicing and cleaning device with expansion pressure for the interior cleaning
of inorganic residues and optimum care.
(no validated cleaning of the interior in accordance with German RKI require-
ments)
KaVo recommends servicing the product as part of the reprocessing after each
use, i.e. after each cleaning, disinfection, and before each sterilisation.
▶
Remove the tool from the medical device.
▶
Service the product in the QUATTROcare.
See also:
2
Instructions for use KaVo QUATTROcare 2104 / 2104A / 2124A
Servicing the chuck
KaVo recommends servicing the chucking system once weekly.
▶
Remove the tool from the medical device.
▶
Plug the spray nipple of the chuck servicing set onto the QUATTROcare plus
Spray.
▶
Position the tip of the spray nipple in the opening, and apply the spray.
▶
Press the spray key for 1 to 2 seconds.
7.5 Packaging
Note
The sterile goods package must be large enough for the product so that the
packaging is not stretched.
The quality and use of the packaging of the items to be sterilised must satisfy
the applicable standards and be appropriate for the sterilisation process!
▶
The medical device must be packed before sterilisation.
27 / 32