Instructions for use elements™ 8:1
7 Reprocessing steps in accordance with ISO 17664 | 7.2 Manual Reprocessing
7.2.4 Manual internal disinfection
WARNING
Incomplete disinfection.
Infection hazard.
▶
Only use disinfection procedures that are verified to be bactericidal, fungi-
cidal and virucidal.
▶
If the disinfectants used do not meet these requirements, the process must
be concluded by disinfection of the unit(s) without packaging using a steam
steriliser.
NOTICE
Never disinfect the medical device with chloride-containing products.
Malfunction and material damage.
▶
Only disinfect in a washer disinfector or manually.
The efficacy of manual internal disinfection must be demonstrated by the man-
ufacturer of the disinfection agent. With KaVo products, use only disinfection
agents that have been released by KaVo with respect to the compatibility of
materials (e.g. WL-cid / made by ALPRO).
▶
Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.
▶
Hold the can vertically.
▶
Press the spray key for at least 3 seconds.
▶
Remove the medical device from the spray attachment and let the disinfect-
ant act for 2 minutes.
▶
Follow the instructions for use of the disinfectant.
7.2.5 Manual drying
Use KaVo DRYspray for subsequent drying of the air, water and gear unit ducts.
▶
Cover the medical device with the KaVo CLEANpac bag, and place it on the
corresponding care adapter.
▶
Hold the can vertically.
▶
Press the spray key for at least 3 seconds.
See also:
2
KaVo DRYspray Instructions for Use
▶
Immediately after drying, lubricate the KaVo medical device with care
agents from the KaVo care system.
See also:
2
7.4 Care products and systems - Servicing, Page 25
23 / 32