F9 Selecting statistics:
The statistics symbol appears on the
display.
1. Put-on sample
(example)
Confirm with
. The value will be
saved to the alibi memory and printed
out if a printer is connected.
(example)
The information line shows “1“ for the
first weighing process.
Remove the sample from the weighing
plate and apply the same sequence of
operations for additional samples.
(example)
After all values have been saved, the
results are ready for evaluation under
statistics.
Press the
F
-key.
The arrow keys allow you to display all
data:
Examples:
N
9
SUM
455.600
G
AVG
50.6222
G
MIN
49.939
G
MAX
51.380
G
DIF
1.441
G
SDV
0.39605
G
RDV
0.78
%
PRINTING
74