7.2.3 P1.3 BE.JUST. User adjustment (model AES-C only)
This function allows the user to select a test weight of an arbitrary weight value.
Having said this, the test weight should not amount to less than 30% of the maximum range.
This function is not accessible in calibrated weighing scales!
Start the process in weighing mode
as shown on the picture.
Digits will appear on the display, with
the digit on the left flashing. Use the
arrow keys to enter your desired
adjustment weight and confirm by
pressing
.
Care should be taken to ensure that
there are no objects on the weighing
plate, then confirm by pressing
.
Wait until the weighed value for the
set adjustment weight appears.
Place the required adjustment
weight carefully in the centre of the
weighing plate and press the
key. The display is showing
CALIBRATION,
adjustment is
starting.
Adjustment is completed when the
message
REMOVE MASS
appears;
remove adjustment weight. After
successful adjustment the weighing
scale will automatically return to sub-
menu item
P 1.3
.
Return to weighing mode by
repeatedly pressing
UNIT
AEJ-C/AES-C/PLJ-C-BA-e-1513