19.5 Alibi memory
The weighing scale software can save up to 100 000 measurements. Measurements
are saved automatically each time the
-key is pressed. In addition to
measurements other data is also saved:
•
Date of measurement
•
Time of measurement
•
Measuring result (weight)
•
Tare value
•
Logged-in user
•
Name of weighed product
•
Measurement one will be deleted as soon as measurement 100 001 is
saved.
•
Measurements saved to the alibi memory cannot be deleted.
1
Number of measurement:
Format: b4.5n N=number of measurement
2
Date of measurement
3
Time of measurement
Each can be invoked as described above (chap. 18.4).
If menu item b4.5.1.7 is selected and a printer connected, data can be printed by
pressing
. (Dependent on settings selected in GLP report chap. 15).
AEJ-C/AES-C/PLJ-C-BA-e-1513
113
Summary of Contents for PLJ-C
Page 10: ...1 1 Abmessungen AEJ C AES C PLJ C 9 AEJ C AES C PLJ C BA d 1513...
Page 20: ...19 AEJ C AES C PLJ C BA d 1513...
Page 121: ......
Page 129: ...1 1 Dimensions AEJ C AES C PLJ C AEJ C AES C PLJ C BA e 1513 9...
Page 139: ...AEJ C AES C PLJ C BA e 1513 19...
Page 241: ......
Page 249: ...1 1 Dimensions AEJ C AES C PLJ C AEJ C AES C PLJ C BA f 1513 9...
Page 259: ...AEJ C AES C PLJ C BA f 1513 19...