13.4 Percent determination
Percent determination allows weight display in percent, in relation to a reference
weight. The reference weight may be determined by weighing or by entering it
numerically.
•
Determining reference by weighing
The function DETERMINE REFERENCE WEIGHT is allocated to the
F1
-key.
Go to menu and select P2.6 DEVIATION
Use the arrow keys to select P2.6.6 SHORTCUTS
Press
F2
, P2.6.6.1 will be displayed along with the previous function for the
F1
-key.
Again press
F2
and the information line will start flashing.
Use the arrow keys to select DETERMINE REFERENCE WEIGHT and
Confirm by
; this function has now been allocated to
F1
.
In weighing mode:
This completes setting the operating
mode weighing by percentage.
The percent symbol will be
displayed at the upper margin of the
display.
The setting SAMPLE 0.0 G will be
displayed at the lower margin.
Put the reference weight equalling
100% on the weighing plate.
(example)
Example: 1 kg is supposed to be
equivalent to 100%.
The reference weight of 1 kg has now been saved
as 100% value.
Remove reference weight and place load. The weight will be displayed in
percent according to reference weight.
AEJ-C/AES-C/PLJ-C-BA-e-1513
69
Summary of Contents for PLJ-C
Page 10: ...1 1 Abmessungen AEJ C AES C PLJ C 9 AEJ C AES C PLJ C BA d 1513...
Page 20: ...19 AEJ C AES C PLJ C BA d 1513...
Page 121: ......
Page 129: ...1 1 Dimensions AEJ C AES C PLJ C AEJ C AES C PLJ C BA e 1513 9...
Page 139: ...AEJ C AES C PLJ C BA e 1513 19...
Page 241: ......
Page 249: ...1 1 Dimensions AEJ C AES C PLJ C AEJ C AES C PLJ C BA f 1513 9...
Page 259: ...AEJ C AES C PLJ C BA f 1513 19...