PLJ-GM-BA-e-1614
13
6.6.3 Sequence of temperature-controlled adjustment:
At a change in temperature of 3°C the balance announces
the internal adjustment via the display „CALib“
The balance starts the adjustment process and displays
„CAL-t“.
If the adjustment process is finished, the balance will
automatically return into the weighing mode.
6.6.4 Sequence of time-controlled adjustment:
After 3 hours, the automatic adjustment is started. „CALib“
is displayed.
The balance starts the adjustment process and displays
„CAL-C“.
If the adjustment process is finished, the balance will
automatically return into the weighing mode.
6.6.5 Manual adjustment with internal weight
In weighing mode press
. „CALib“ appears.
During the adjustment process „CAL-H“ is displayed
If the adjustment process is finished, the balance will
automatically return into the weighing mode.
Use the
button to cancel the adjustment process at any
time.
is briefly displayed.
Then the balance returns automatically into weighing mode.
TARE
F