![KERN PLJ 600-2GM Operating Instruction Download Page 110](http://html1.mh-extra.com/html/kern/plj-600-2gm/plj-600-2gm_operating-instruction_1957071110.webp)
PLJ-GM-BA-e-1614
52
9.5.3 Continuous output
The balance may be operated in a mode enabling continuous output of weighing
results. This mode can be turned on/off by commands via RS232.
Data record format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S
I
Space
Stability
indicator
Space
Pre
sign
Weig
ht
Spac
e
Unit
CR
LF
Stability indicator Space if stable,
? if not stable
^ if overload
v if underload
Presign:
Space, if positive
negative sign, if negative
Weight:
9 signs, right justified
Unit:
3 signs, left justified
10 Error messages
„Err2“:
Value outside zero range
„Err3“:
Value outside taring range
„Err4“:
Calibration weight outside allowable range (+-1% for
calibration weight)
„Err5“:
Piece weight smaller than readability
„Err7“:
Disconnection time was too short
(should be more than 3 seconds)
„Err8“:
Impossible to carry out taring / resetting
„NULL“:
Underload
„FULL2“:
Weighing range exceeded
„LH“:
Initial weight error Weight of weighing plate outside
allowable tolerance of 10%
„5-Full“:
Display range exceeded upon totalising