PLJ-GM-BA-e-1614
12
6.5.2 Balance zero display
If an exact zero reading is not displayed on the balance in spite of the weighing dish
being empty, press the
and the balance will start resetting to zero
[
a
] .
To
reset the balance to zero, the
may be used if the load amounts to less than 2%
of the maximum weight. If the load exceeds 2% error message
ERR2
will be
displayed.
6.6 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
6.6.1 Automatic adjustment with internal weight
With the internal adjustment weight, the weighing accuracy can be checked and re-
adjusted at any time.
The automatic adjustment function is always enabled. You can start adjustment at
any time by pressing the
CAL
-key manually.
Automatic adjustment is started each time
•
after the weighing balance was disconnected from the mains
•
when temperature changes take place
•
at the end of a time interval
6.6.2 Sequence of adjustment after the balance having been switched on:
After the balance having performed a self check, automatic
adjustment will start and in the display appears „CALib“
During the adjustment process „CAL-A“ is displayed
If the adjustment process is finished, the balance will
automatically return into weighing mode.
TARE