6.2.1 Setup
Level balance with foot screws until the air bubble of the
water balance is in the prescribed circle.
6.2.2 Scope of delivery
Serial accessories:
•
Balance
•
Weighing plate
•
Mains power supply
•
Operating Manual
•
Protective cover
•
Hook for underfloor weighing (only model PLJ 6000-1GM)
6.3 Mains connection
Power is supplied via the external mains adapter. The stated voltage value must be
the same as the local voltage. Only use original KERN mains adapters. Using other
makes requires consent by KERN.
6.4 Connection of peripheral devices
Before connecting or disconnecting of additional devices (printer, PC) to the data
interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they
are ideally tuned to your balance.
6.5 Initial Commissioning
In order to obtain exact results with the electronic balances, your balance must have
reached the operating temperature (see warming up time chap. 1).
During this warming up time the balance must be connected to the power supply
(mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.
6.5.1 Stability display
If the display shows the stability display
[
]
the balance is in a stable status. If the
status is instable the [
] display disappears.