3.
Error Messages
Fault messages
Error
number
Description
Error of control sum
1.1
Error data transfer
Error A/D
1.2
Error A/D converter
Overcrossed range
2.1
Outside maximum range
Overcrossed range
2.2
Outside maximum range
A/D Null
2.3
Error A/D converter
A/D Full
2.4
Error A/D converter
Tare/zero above the range
2.5
Outside of weighing range
Tare above the range
2.6
Outside of taring range
Result > 10% Max
2.7
Result > 10 %
Result > 4% Max
2.8
Load on weighing plate too heavy
Difference > 1% Max
2.9
Difference between cal. weight
stored/current cal. weight > 1%
Sample mass < 20 mg
2.10
Sample < 20 mg
Sample mass above the
range
2.11
Sample out of set range
Above the range
3.1
Value of parameter outside
Faulty value
3.2
Value of parameter outside
Incorrect password
3.4
Wrong password
Error of notice
4.1
Parity error
4.2
Table error
4.3
Suspended transmission
CTS
4.4
Suspended transmission
XOFF
4.5
Error data transfer
Incorrect date
5.1
Wrong data
Overcrossed time
6.1
Timeout
MLS…N-SH-e-0812
5