Printout of Factory Parameters
Open the housing and set jumper „
JP1
“ on main board 172xxx.PCB to access the
service menu.
Connect a standard printer to RS 232 data interface of the moisture analyzer.
Press the ON/OFF key (main switch) to turn on the balance.
Enter main menu
Set index next to parameter
P7 Globals
P1 alibration
P2 GLP
P3 D
P5 RS-232
P6
P7
C
ate/Time
P4 Redout
Printouts
Globals
Setup
29.12.04
Enter menu
P7 Globals
Enter submenu
P7 09 Par. printout
To get a printout of the factory
parameters
P0 01
to
P0 75
set
Start= 0.01
Stop= 0.75
Confirm your setup with PRINT
Press PRINT to start the printout