P0 17
Temp. corr. Starts temp. compensation in a room with
regulated temperature (range ± 1°C). Correction is
done in temp. 18°C up to 28°C.
The parameters/values are stored after doing
temperature compensation (parameters P0 27, P0
28).
P0 18
Start div
Preload (with pan) - balance find it automatically
P0 19
Cal. factor
Calibration factor during factory calibration
P0 20
Start div f. Preload after correction
P0 21
Cal. factor f.
Not documented
P0 22
T. max
Not documented
P0 23
Start T
Not documented
P0 24
Factor T
The factor during the temperature compensation
P0 25
Start f. T
Not documented
P0 26
Factor f. T
The factor during temperature compensation after
filtering
P0 27
1’st point T
First temperature correction set by 18°C
P0 28
Start T 2
Not documented
P0 29
Factor T 2
Factor during temperature compensation 2’nd point
P0 30
Start f.T2
Not documented
P0 31
Factor f. T 2
The factor during temperature compensation after
filtering 2’nd temperature point
P0 32
2’nd point T
Second temperature correction set by 28°C
P0 33
Temp. Z factor
Temperature compensation factor of zero
P0 34
Temp. S factor
Temperature compensation factor of sensitivity
P0 35
Lin. points
Lin. value 1
Lin. factor 1
Lin. value 2
Lin. factor 2
Lin. value 3
Lin. factor 3
.
.
.
.
.
.
.
.
P0 64
Lin. value 15
P0 65
Lin. factor 15
P0 66
Lin. value 16
P0 67
Lin. factor 16
Those parameters are stored after doing linearity
correction. These parameters (the stored values)
can be changed manually (max. 16 points).
Here you can change the linearity factor manually
also
without
doing the linearity correction by using
external weights.
P0 68
Lin. factor A
Parabolic linearity factor
P0 69
Lin. factor B
Parabolic linearity factor
P0 70
Up/Down par.
Printout of factory parameters, printout of moisture
analyzer parameters or receipt of moisture
analyzer parameters
P0 71
Stack info.
Information about the stored settings when the
moisture analyzer will be started
P0 72
Bootloader
Select this parameter for software download /
update
8