4.2
Description of factory parameters
Press the ON/OFF key (main switch) to turn on the balance
Enter main menu
Set index next to parameter
P0 Factory
Enter submenu
P0 Factory
Parameter
number
Parameter name
Description
P0 01
Factory deff
Delete all settings in balance and load program
primal settings
» Factory setting (Reset) /
PLEASE NOTICE you
have to set all parameters new!!!
P0 02
Balance Id
Change factory number (serial number)
P0 03
Full scale
(scale capacity)
Scale capacity (max) = weighing range +9
divisions: e.g. max range
50,09g
P0 04
Div
Reading precision (resolution), e.g. 0.001
P0 05
Ext. cal. mass
Value of external calibration weight [
g
], e.g. 50.0g
P0 06
Autozero range
Range of autozero function (possibility to change
value from 0.1d to 10.0d) Normal
12= 3.0 d
P0 07
Autozero delay
Time of autozero (possibility to change value from
0.2s
to
3.0s
) Normal
6= 2.0 s
P0 08
Stable range
Stable range of measuring results (possibility to
change value from 0.1d to 10.0d)
Normal
11= 2.0 d
P0 09
Stable speed
Time to define stable measuring results (possibility
to change value from 0.2s to 3.0s)
Normal
6= 2.0 s
P0 10
Filter range
Digital filter (possibility to change value from 1d to
10000d)
Normal
5 – 20 d
P0 11
Show A/D div
Value of A/D converter (of the load)
P0 12
Show A/D T1
A/D value of temperature sensor in the coil
P0 13
Show A/D T2
A/D value of temperature sensor in the drying
chamber
P0 14
Factory cal. Starts factory calibration with external calibration
weight
P0 15
Factory T cal.
Starts calibration of the temperature sensor inside
the drying chamber
P0 16
Linear. corr.
Starts linearity procedure
with
external weights.
You can inscribe max. 16 points of linearity.
MLS…N-SH-e-0812