PRELIMINARY OPERATIONAL CHECK. 

 A simple operational check after unpacking and

before equipment installation is advisable to ascertain whether the power supply has suffered damage
resulting from shipping. 

NOTE: This test must be performed with all I/O ports disconnected, and ANALOG I/O SETTINGS

switch positions 3, 4 and 5 set to OFF (up).

1 With POWER circuit breaker set to OFF position, connect the power supply to source power (see

See “Input Connections.” on page 5.).

2 With no load connected, set POWER circuit breaker to the ON position. Each time the unit is turned

on an internal self-test is performed (see “Turning The Power Supply On.” on page 7). After the test

has been successfully completed, the status display reads SET (setpoint mode), the DC VOLTS dis-

play reads 0 Volts and the DC AMPERES display reads minimum Amperes (see Note 1, below).

NOTES: 1. A minimum programmed current (actual value depends on model) is required to ensure proper

operation of the power supply under all load conditions. Programmed current is automatically set

to be at least the minimum current.

2. If an error indication is blinking in the Status display, refer to User Manual for an explanation of

error codes.

3 Rotate VOLTAGE adjust knob clockwise. Verify that DC VOLTS display increases in large (X 100)

increments. 

4 Press and rotate VOLTAGE adjust knob clockwise. Verify the DC VOLTS display increases in finer

increments than step 3, then release knob. 

5 Adjust VOLTAGE adjust knob clockwise until Status display reads >MAX. Tap either VOLTAGE or

CURRENT adjust knob once to enter values. Verify Status display is blank

6 Connect a digital voltmeter (DVM) to the (M+) and (M–) terminals on the rear panel.
7 Press and release DC OUTPUT switch to enable the output. Verify DC OUTPUT indicator lights.
8 Compare the programmed output voltage value (step 5) with the voltage reading of the DVM; the dif-

ference between the two should not exceed 0.05% of the maximum voltage of the unit.

9 Compare the voltage reading of the DC VOLTS display with that of the DVM; the difference between

the two should not exceed 0.1% of the maximum voltage of the unit.

10 Enter different value for output voltage, then repeat steps 4 and 5 using different values for pro-

grammed voltage.

11 Disable the output by pressing and releasing DC OUTPUT switch; verify front panel DC VOLTS and

DC AMPERES displays read 0.0V and minimal current and the DC OUTPUT indicator is off.

12 Set POWER ON/OFF circuit breaker to OFF and disconnect from source power, load and test equip-

ment.

Summary of Contents for KLP 10-150

Page 1: ...as a handy reference Refer to the KLP Developer s Guide for a full description of the digital interfaces their associated drivers and the SCPI command language Firmware Version Refer to www kepcopowe...

Page 2: ...he PRELIMINARY OPERATIONAL CHECK If any indication of damage is found file an imme diate claim with the responsible transport service EQUIPMENT SUPPLIED See Table 2 TABLE 1 MODEL PARAMETERS Model numb...

Page 3: ...2 0381 IEEE 488 GPIB Cable 1m long Connect KLP Power Supply to GPIB bus SNC 488 1 IEEE 488 GPIB Cable 2m long Connect KLP Power Supply to GPIB bus SNC 488 2 IEEE 488 GPIB Cable 4m long Connect KLP Pow...

Page 4: ...n error indication is blinking in the Status display refer to User Manual for an explanation of error codes 3 Rotate VOLTAGE adjust knob clockwise Verify that DC VOLTS display increases in large X 100...

Page 5: ...ords and a user wired source power connector are available as accessories see Table 3 Plug the source power connector into the source power inlet connector at the rear panel LOAD CONNECTIONS Connect t...

Page 6: ...S see Figure 2 FIGURE 2 KLP SERIES REAR PANEL VIEW JUMPERS INSTALLED FOR LOCAL SENSING REMOTE SENSING SELECT First remove the factory installed local sensing jumpers between S and M and between M and...

Page 7: ...e of the internal relay operation via the LAN interface or analog signals and setting coarse fine adjustment preference of the VOLTAGE and CURRENT controls An Installation Operation Summary is also in...

Page 8: ...how the actual output voltage and current Before enabling the output ALWAYS check the set points to avoid possible damage to the load 1 Tap either the VOLTAGE or CURRENT control The status display rea...

Page 9: ...s showing actual output current Similarly rotating the VOLTAGE control while in CC mode affects the voltage limit To change the limits to a precise value refer to SETPOINT ADJUSTMENT Voltage and curre...

Page 10: ...he output off flashes an overvoltage OVP or overcurrent OCP protection error message on the status display and sets a status bit that can be retrieved through the RS 232 GPIB or LAN port The unit must...

Page 11: ...13 228 1616 REV 7 11 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 4 KLP OUTLINE DIMENSIONS SHEET 1 O...

Page 12: ...28 1616 REV 7 070113 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 4 KLP OUTLINE DIMENSIONS SHEET 2 O...

Reviews: