ENABLING/DISABLING OUTPUT POWER. 

When the power supply is turned on, the output is auto-

matically disabled (DC OUTPUT LED is off), the DC VOLTS and DC AMPERES displays show the pro-
grammed output voltage and current set points and “SET” appears in the status display. 

1 To enable the output, first exit SET mode by tapping either the VOLTAGE or CURRENT control, then

press and release the DC OUTPUT switch. The associated green LED lights to indicate output

power is applied to the load. Each time you exit SET mode, the setpoint values are stored for possi-

ble recall if the power supply is turned off or the output is disabled.

2 To disable the output, press and release the DC OUTPUT switch again. The DC OUTPUT indicator

goes off. 

NOTE: While in the setpoint mode (Status display reads SET) the output cannot be turned on

(pressing DC OUTPUT has no effect). If output was on while setpoint mode was entered,
pressing DC OUTPUT will disable the output.

CHECKING OR CHANGING VOLTAGE/CURRENT SETPOINTS. 

CAUTION: When the output is disabled, the DC VOLTS and DC AMPERES displays show the

actual output voltage and current. Before enabling the output ALWAYS check the set-
points to avoid possible damage to the load.

1 Tap either the VOLTAGE or CURRENT control. The status display reads SET, and the DC VOLTS

and DC AMPERES displays show the stored setpoints. 

2 To accept the displayed value, tap the associated adjustment control again. To change the value,

rotate the control (press the control in while rotating for fine adjustment; or see User Manual to set

coarse/fine preference), then tap the adjustment control again to accept the new setting. 

NOTE: Before changing the setpoint, note the displayed setpoint. If you decide not to change the

value after rotating the control, you must rotate the control to the value noted, then tap the
adjustment control to accept. 

VIRTUAL MODEL SETTING. 

The virtual model establishes a maximum programmable voltage and

current for the unit within the 1200W power limitation and the maximum voltage and current ratings listed
in Table 1. The unit will not accept programmed values beyond these values. 

1 Using a thin tool (e.g., a paper clip), press the FUNCTION switch until the status display reads VIRT.

The DC VOLTS and DC AMPERES displays show the programmed maximum voltage and current of

the virtual model. 

2 To exit, rotate either VOLTAGE or CURRENT control until Status display reads EXIT, then tap DC

OUTPUT switch. To change the virtual model limits refer to the User Manual.

Summary of Contents for KLP 10-150

Page 1: ...as a handy reference Refer to the KLP Developer s Guide for a full description of the digital interfaces their associated drivers and the SCPI command language Firmware Version Refer to www kepcopowe...

Page 2: ...he PRELIMINARY OPERATIONAL CHECK If any indication of damage is found file an imme diate claim with the responsible transport service EQUIPMENT SUPPLIED See Table 2 TABLE 1 MODEL PARAMETERS Model numb...

Page 3: ...2 0381 IEEE 488 GPIB Cable 1m long Connect KLP Power Supply to GPIB bus SNC 488 1 IEEE 488 GPIB Cable 2m long Connect KLP Power Supply to GPIB bus SNC 488 2 IEEE 488 GPIB Cable 4m long Connect KLP Pow...

Page 4: ...n error indication is blinking in the Status display refer to User Manual for an explanation of error codes 3 Rotate VOLTAGE adjust knob clockwise Verify that DC VOLTS display increases in large X 100...

Page 5: ...ords and a user wired source power connector are available as accessories see Table 3 Plug the source power connector into the source power inlet connector at the rear panel LOAD CONNECTIONS Connect t...

Page 6: ...S see Figure 2 FIGURE 2 KLP SERIES REAR PANEL VIEW JUMPERS INSTALLED FOR LOCAL SENSING REMOTE SENSING SELECT First remove the factory installed local sensing jumpers between S and M and between M and...

Page 7: ...e of the internal relay operation via the LAN interface or analog signals and setting coarse fine adjustment preference of the VOLTAGE and CURRENT controls An Installation Operation Summary is also in...

Page 8: ...how the actual output voltage and current Before enabling the output ALWAYS check the set points to avoid possible damage to the load 1 Tap either the VOLTAGE or CURRENT control The status display rea...

Page 9: ...s showing actual output current Similarly rotating the VOLTAGE control while in CC mode affects the voltage limit To change the limits to a precise value refer to SETPOINT ADJUSTMENT Voltage and curre...

Page 10: ...he output off flashes an overvoltage OVP or overcurrent OCP protection error message on the status display and sets a status bit that can be retrieved through the RS 232 GPIB or LAN port The unit must...

Page 11: ...13 228 1616 REV 7 11 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 4 KLP OUTLINE DIMENSIONS SHEET 1 O...

Page 12: ...28 1616 REV 7 070113 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 4 KLP OUTLINE DIMENSIONS SHEET 2 O...