DESCRIPTION. 

The KLP Power Supply Series are universal input, automatic crossover, 1200-watt

constant power, voltage/current stabilizers with a full rectangular output characteristic. There are seven
standard models (also referred to as -1200) and seven E -Series models (formerly referred to as -1.2K)
(see Table 1). The standard models are GPIB and RS 232 compatible; the E-Series models are GPIB
and LAN compatible; all models allow remote analog programming. For each model a multitude of virtual
models can be configured. The “rated voltage, maximum current at rated voltage” and “rated current and
maximum voltage at rated current” parameters define the virtual models available within the limit of 1200
Watts of output power.

OPTIONS. 

The Rapid Output Discharge Circuit (RODC) option (suffix R) is available on all KLP mod-

els. This circuit rapidly discharges the output capacitance, thus significantly reducing response time to
reductions in output voltage (see KLP User Manual for details).

UNPACKING. 

This instrument has been thoroughly inspected and tested prior to packing and is

ready for operation. After careful unpacking, inspect for shipping damage before attempting to operate.
Perform the PRELIMINARY OPERATIONAL CHECK. If any indication of damage is found, file an imme-
diate claim with the responsible transport service.

EQUIPMENT SUPPLIED

.

  

See Table 2.

TABLE 1.  MODEL PARAMETERS

Model number

Rated 

Voltage

Range

(1)

Maximum 

current for 

rated voltage

Minimum

programmable 

current 

Rated 

Current

Range

(1)

Maximum 

voltage for 

rated current

Ripple 

and 

noise
p-p

(2)

Efficiency

@115 Va-c

KLP 10-150

(3)

0-10V

120A@10V

1.9A

0-150A

8V@150A

60 mV

80%

KLP 20-120

(3)

0-20V

60A@20V

1.5A

0-120A

10V@120A

60 mV

82%

KLP 36-60

(3)

0-36V

33.3A@36V

0.8A

0-60A

20V@60A

60 mV

83%

KLP 75-33

(3)

0-75V

16A@75V

0.4A

0-33.3A

[email protected]

60 mV

84%

KLP 150-16

(3)

0-150V

8A@150V

0.2A

0-16A

75V@16A

125 mV

86%

KLP 300-8

(3)

0-300V

4A@300V

0.1A

0-8A

150V@8A

150 mV

87%

KLP 600-4

(3)

0-600V

2A@600V

0.05A

0-4A

300V@4A

150 mV

88%

(1) The maximum current and voltage are constrained by the 1200 watt power limitation.
(2) Bandwidth: 20MHz; low frequency ripple may be higher at loads less than 30 Watts.
(3) Specifications listed apply to both standard (-1200) and E-Series (suffix E) models.

TABLE 2.  EQUIPMENT SUPPLIED

ITEM

PART NUMBER

QUANTITY

Jumper (24 AWG or larger bus wire) for local sensing connections 

172-0638

2

Analog I/O port mating connector

142-0528

1

Nut, 1/4-20, with captive lockwasher (on output studs of 10V, 20V and 36V models only)

102-0175

2

Quick Start Guide *

228-1616

1

* User Manual and Developer’s Guide are available as free downloads from:
       www.kepcopower.com/support/opmanls.htm#klp
  Drivers are available as free downloads from
       www.kepcopower.com/drivers/drivers-dl3.htm#klp

Summary of Contents for KLP 10-150

Page 1: ...as a handy reference Refer to the KLP Developer s Guide for a full description of the digital interfaces their associated drivers and the SCPI command language Firmware Version Refer to www kepcopowe...

Page 2: ...he PRELIMINARY OPERATIONAL CHECK If any indication of damage is found file an imme diate claim with the responsible transport service EQUIPMENT SUPPLIED See Table 2 TABLE 1 MODEL PARAMETERS Model numb...

Page 3: ...2 0381 IEEE 488 GPIB Cable 1m long Connect KLP Power Supply to GPIB bus SNC 488 1 IEEE 488 GPIB Cable 2m long Connect KLP Power Supply to GPIB bus SNC 488 2 IEEE 488 GPIB Cable 4m long Connect KLP Pow...

Page 4: ...n error indication is blinking in the Status display refer to User Manual for an explanation of error codes 3 Rotate VOLTAGE adjust knob clockwise Verify that DC VOLTS display increases in large X 100...

Page 5: ...ords and a user wired source power connector are available as accessories see Table 3 Plug the source power connector into the source power inlet connector at the rear panel LOAD CONNECTIONS Connect t...

Page 6: ...S see Figure 2 FIGURE 2 KLP SERIES REAR PANEL VIEW JUMPERS INSTALLED FOR LOCAL SENSING REMOTE SENSING SELECT First remove the factory installed local sensing jumpers between S and M and between M and...

Page 7: ...e of the internal relay operation via the LAN interface or analog signals and setting coarse fine adjustment preference of the VOLTAGE and CURRENT controls An Installation Operation Summary is also in...

Page 8: ...how the actual output voltage and current Before enabling the output ALWAYS check the set points to avoid possible damage to the load 1 Tap either the VOLTAGE or CURRENT control The status display rea...

Page 9: ...s showing actual output current Similarly rotating the VOLTAGE control while in CC mode affects the voltage limit To change the limits to a precise value refer to SETPOINT ADJUSTMENT Voltage and curre...

Page 10: ...he output off flashes an overvoltage OVP or overcurrent OCP protection error message on the status display and sets a status bit that can be retrieved through the RS 232 GPIB or LAN port The unit must...

Page 11: ...13 228 1616 REV 7 11 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 4 KLP OUTLINE DIMENSIONS SHEET 1 O...

Page 12: ...28 1616 REV 7 070113 KEPCO INC 131 38 SANFORD AVENUE FLUSHING NY 11355 U S A TEL 718 461 7000 FAX 718 767 1102 http www kepcopower com email hq kepcopower com FIGURE 4 KLP OUTLINE DIMENSIONS SHEET 2 O...

Reviews: