KLP070312
3-21
3.2.16.5 ANALOG INPUT FULL SCALE CALIBRATION
Enter
UTIL
menu (see PAR. 3.2.16, steps 1 through 3) and rotate either
VOLTAGE
or
CUR-
RENT
control until status display reads
ANPR
.
• The
DC VOLTS
display shows the analog input voltage value that produces the full scale
output voltage established by the virtual model.
• The
DC AMPS
display shows the analog input voltage value that produces the full scale
output current established by the virtual model.
1. To change the settings tap the
DC OUTPUT
switch. Note that
DC VOLTS
and
DC
AMPERES
displays begin blinking.
2. Rotate
VOLTAGE
control to set the analog voltage value required to produce the full scale
output voltage.
3. Rotate the
CURRENT
control to set the analog voltage value required to produce the full
scale output current.
4. Tap the
DC OUTPUT
switch to accept the new settings (blinking stops) or use a thin tool
(e.g., a paper clip) to press the
FUNCTION
switch to cancel the operation.
5. To exit, rotate either
VOLTAGE
or
CURRENT
control until the status display reads
RTN
, then
tap the
DC OUTPUT
switch. To exit the
UTIL
menu rotate either
VOLTAGE
or
CURRENT
control until the status display reads
EXIT
, then tap the
DC OUTPUT
switch.
3.2.16.6 CALIBRATION
Refer to PAR 4.3.
3.3
REMOTE MODE PROGRAMMING USING DIGITAL INTERFACES
For digital remote programming refer to the KLP Developer’s Guide.
3.3.1
FACTORY DEFAULT SETTINGS
3.3.1.1
GPIB FACTORY DEFAULTS
The factory default GPIB address is 6; refer to PAR. 3.2.11 to change the GPIB address.
3.3.1.2
RS 232 FACTORY DEFAULTS [STANDARD MODELS ONLY]
The factory default RS 232 settings are as follows; refer to PAR. 3.2.12 to change the RS 232
baud rate (standard models only):
• Baud rate: 38400
• Parity:
None
• Data Bits
8
• Stop Bits
1
• Echo
OFF
• XON
ON (enabled) NOTE: Kepco strongly recommends the XON XOFF method
for data transfer via RS 232 protocol for all Kepco products. If this method is
not selected, it is the user's responsibility to ensure completion of any
response by the power supply prior to issuance of subsequent commands.
Summary of Contents for KLP 10-150-2
Page 2: ......
Page 6: ......
Page 12: ...vi KLP070312 FIGURE 1 1 KLP SERIES POWER SUPPLY...
Page 19: ...KLP 070312 1 7 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 1 OF 2...
Page 20: ...1 8 KLP 070312 FIGURE 1 3 KLP SERIES POWER SUPPLY MECHANICAL OUTLINE DRAWING SHEET 2 OF 2...
Page 27: ...KLP HV 070312 2 3 FIGURE 2 2 KLP SERIES REAR PANEL SWITCH AND CONNECTORS...
Page 40: ......
Page 78: ......