KEOFITT W9 USER MANUAL V.4
PAGE 28
6.7 Flow
The graphs below illustrate (for water at 20°C/68°F) the following:
• Pressure drop across valve
*
) as a function of the flow for different positions of the turn knob
• Pressure drop for flow between the inlet and outlet ports (CIP ports)
*
)From tank/pipe side to lower port (outlet) with upper port blocked.
Based on the tank pressure and the requested sample flow the graphs may be used to get an indication
of to which degree the valve must be opened.
The generally accepted sampling time is around 10 sec. for small samples and around 30 sec. for larger
samples. As usual sample sizes are between 100 ml and 1000 ml the needed flow lies from 600 to 2000
ml/min.
As the pressure on the sample side usually is 0 bar(g) the pressure drop across the valve equals the pro-
cess pressure (tank pressure or line pressure).
The volume flow through a valve is given by:
k
v
= Q
1000x∆p
————
ρ
Symbol
Unit
Description
�
k
v
m
3
/h
Flow in m
3
/h through a valve at a pressure drop of 1 bar as defined in VDE/VDI
norm 2173.
�
Q
m
3
/h
Volume flow through the valve
ρ
kg/dm
3
Density of the fluid. For Water it is 1.
Δp
bar
Pressure drop across valve.
As the gauge pressure at the valve outlet usually is 0 bar(g) the pressure drop is
often equal to the gauge pressure at the input (the process side)
Summary of Contents for W9
Page 1: ...DON T GAMBLE WITH YOUR SAMPLE W9 SAMPLING VALVE User Manual ...
Page 4: ...KEOFITT W9 USER MANUAL V 4 PAGE 4 ...
Page 44: ...KEOFITT W9 USER MANUAL V 4 PAGE 44 16 MEMBRANES 16 1 Silicone membrane art no 600051 ...
Page 45: ...KEOFITT W9 USER MANUAL V 4 PAGE 45 16 2 Silicone membrane for Micro Port art no 600251 ...
Page 46: ...KEOFITT W9 USER MANUAL V 4 PAGE 46 16 3 EPDM membrane art no 600052 ...
Page 47: ...KEOFITT W9 USER MANUAL V 4 PAGE 47 16 4 EPDM membrane for Micro Port art no 600252 ...
Page 48: ...KEOFITT W9 USER MANUAL V 4 PAGE 48 16 5 FFKM membrane art no 600053 ...
Page 49: ...KEOFITT W9 USER MANUAL V 4 PAGE 49 16 6 PTFE membrane art no 850055 ...