KEOFITT W9 USER MANUAL V.4
PAGE 21
5.2 Chemical disinfection
Immediately following the CIP, perform the disinfection, if required. The disinfection takes place with the
valve remaining in its closed position.
There are 2 recommended ways to carry out the disinfection:
A) by letting the disinfectant flow through the valve chamber
B) by filling the valve chamber with the disinfectant (advantage: smaller volume of disinfectant
needed and quicker and more reliable disinfection)
Steps to perform, when adopting A:
1. Connect a hose with an empty bottle to the lower hose piece. This
bottle or similar recipient is to collect the disinfectant (step 3) and the
rinsing water (step 6).
2. Fill a flexible bottle with the defined amount of disinfectant.
3. Connect the flexible bottle via a hose to the upper hose piece and
press the disinfectant slowly through the valve to wet the interior of
the valve.
4. Allow the disinfectant to act for the prescribed time.
5. Disconnect the hose from the upper hose piece and connect a flexible
bottle with cleaned water to the upper hose piece.
6. Rinse through the upper hose piece by squeezing the bottle, thus
pressing the water through the valve chamber.
7. Leave the squeezed bottle connected to the hose piece and clamp
the hose to avoid contamination from air being sucked in through the
valve.
Summary of Contents for W9
Page 1: ...DON T GAMBLE WITH YOUR SAMPLE W9 SAMPLING VALVE User Manual ...
Page 4: ...KEOFITT W9 USER MANUAL V 4 PAGE 4 ...
Page 44: ...KEOFITT W9 USER MANUAL V 4 PAGE 44 16 MEMBRANES 16 1 Silicone membrane art no 600051 ...
Page 45: ...KEOFITT W9 USER MANUAL V 4 PAGE 45 16 2 Silicone membrane for Micro Port art no 600251 ...
Page 46: ...KEOFITT W9 USER MANUAL V 4 PAGE 46 16 3 EPDM membrane art no 600052 ...
Page 47: ...KEOFITT W9 USER MANUAL V 4 PAGE 47 16 4 EPDM membrane for Micro Port art no 600252 ...
Page 48: ...KEOFITT W9 USER MANUAL V 4 PAGE 48 16 5 FFKM membrane art no 600053 ...
Page 49: ...KEOFITT W9 USER MANUAL V 4 PAGE 49 16 6 PTFE membrane art no 850055 ...