5.1 Chemical CIP
The CIP takes place with the valve remaining in its closed position.
Perform the following steps:
1. Remove the plugs.
(In the case of a valve with clamp connections there are no plugs
supplied.)
2. Connect a water hose to the upper hose piece.
3. Connect a hose to the lower hose piece and let the hose go to a drain.
4. Flush with clean water.
5. Remove the water hose and let the CIP liquid flow through the upper
hose piece. If the CIP liquid must not go to drain, circulate it or collect
it in a suitable container and dispose of correctly.
6. Reconnect the water hose to the upper hose piece and rinse with
clean water.
If disinfection is not needed the valve is now ready for taking a sample. If disinfection is required proceed
with the steps mentioned in the section “Chemical disinfection” below.
Flush with clean water after sampling. If the process media is sticky, viscous or aggressive or for any
other appropriate reason, do repeat the full CIP cycle after sampling.
WARNING
• Carefully follow the guidelines given for the chemicals involved
• Always remember to use safety goggles when steaming, CIPping, taking samples and all other
operations of the sampling valve
Summary of Contents for W9
Page 1: ...DON T GAMBLE WITH YOUR SAMPLE W9 SAMPLING VALVE User Manual ...
Page 4: ...KEOFITT W9 USER MANUAL V 4 PAGE 4 ...
Page 44: ...KEOFITT W9 USER MANUAL V 4 PAGE 44 16 MEMBRANES 16 1 Silicone membrane art no 600051 ...
Page 45: ...KEOFITT W9 USER MANUAL V 4 PAGE 45 16 2 Silicone membrane for Micro Port art no 600251 ...
Page 46: ...KEOFITT W9 USER MANUAL V 4 PAGE 46 16 3 EPDM membrane art no 600052 ...
Page 47: ...KEOFITT W9 USER MANUAL V 4 PAGE 47 16 4 EPDM membrane for Micro Port art no 600252 ...
Page 48: ...KEOFITT W9 USER MANUAL V 4 PAGE 48 16 5 FFKM membrane art no 600053 ...
Page 49: ...KEOFITT W9 USER MANUAL V 4 PAGE 49 16 6 PTFE membrane art no 850055 ...