1-108 (No.RA054<Rev.001>)
[Analog
Narrow]
1) Adj item:
[An DQT]
Adjust:[****]
2) Adj item:
[L1]
→
[L3]
→
[L5]
→
[C1]
→
[C3]
→
[C5]
→
[H2]
→
[H4]
→
[H6]
Adjust:[****]
PTT : ON
Press [ ]
button to store
the adjustment
value.
1) Adj item:
[DQT Deviation
(Analog Narrow)]
Deviation meter
LPF: 3kHz
HPF: OFF
2) Adj item:
[Low1]
→
[Low3]
→
[Low5]
→
[Center1]
→
[Center3]
→
[Center5]
→
[High2]
→
[High4]
→
[High6]
Press [Transmit]
button.
Press [Apply]
button to store the
adjustment value.
Deviation
meter
Oscilloscope
Panel ANT
Panel [Panel
tuning
mode]
[
],
[
]
[PC
test
mode]
[+], [-]
Write the value as
followings.
K:414(Reference
value)
K2:426(Reference
value)
0.35kHz±0.05
kHz
12. LTR
Deviation
[Analog
Wide]
1) Adj item:
[Aw LTR]
Adjust:[****]
2) Adj item:
[L1]
→
[L3]
→
[L5]
→
[C1]
→
[C3]
→
[C5]
→
[H2]
→
[H4]
→
[H6]
Adjust:[****]
PTT : ON
Press [ ]
button to store
the adjustment
value.
1) Adj item:
[LTR Deviation
(Analog Wide)]
Deviation meter
LPF: 3kHz
HPF: OFF
2) Adj item:
[Low1]
→
[Low3]
→
[Low5]
→
[Center1]
→
[Center3]
→
[Center5]
→
[High2]
→
[High4]
→
[High6]
Press [Transmit]
button.
Press [Apply]
button to store the
adjustment value.
Deviation
meter
Oscilloscope
Panel ANT
Panel [Panel
tuning
mode]
[
],
[
]
[PC
test
mode]
[+], [-]
Write the value as
followings.
K:508(Reference
value)
K2:518(Reference
value)
1.00kHz±0.05
kHz
Item
Condition
Measurement
Adjustment
Specifications
/Remarks
Panel tuning
mode
PC test mode
Test-
equipment
Unit Terminal Unit
Parts
Method