TK-5810
(
BG
)
/5810H
(
BG
)
70
ADJUSTMENT
Item
Condition
Measurement
Adjustment
Specifi cations /
Remarks
Panel tuning mode
PC test mode
Test-
equipment
Unit
Terminal
Unit
Parts
Method
2. AF level
setting
[Panel test mode]
1) CH-Sig: 1-1
Filter:
W
SSG
output
: –47dBm (1mV)
(MOD: 1kHz/±3kHz)
[PC test mode]
1) Channel: 1
Signaling:
1
Bandwidth
: 5kHz/25.0kHz
SSG
output
: –47dBm (1mV)
(MOD: 1kHz/±3kHz)
SSG
DVM
AF VTVM
Dummy
load (4
Ω
)
Rear
panel
ANT
Ext.SP
connector
Volume
knob
Turn the Volume
knob to obtain 2.0V
AF output.
(2W @4
Ω
load)
2.0V±0.3V
3. Sensitivity
check
(Wide)
[Panel test mode]
1) CH-Sig: 1-1
Filter:
W
SSG
output
: –117dBm (0.32µV)
(MOD: 1kHz/±3kHz)
2) CH-Sig: 2-1
3) CH-Sig: 3-1
[PC test mode]
1) Channel: 1
Signaling:
1
Bandwidth
: 5.0kHz/25.0kHz
SSG
output
: –117dBm (0.32µV)
(MOD: 1kHz/±3kHz)
2) Channel: 2
3) Channel: 3
SSG
AF VTVM
Oscillo-
scope
ANT
Ext.SP
connector
Check
12dB SINAD or
more
(Narrow)
1) CH-Sig: 1-1
Filter:
N
SSG
output
: –117dBm (0.32µV)
(MOD: 1kHz/±1.5kHz)
2) CH-Sig: 2-1
3) CH-Sig: 3-1
1) Channel: 1
Signaling:
1
Bandwidth
: 2.5kHz/12.5kHz
SSG
output
: –117dBm (0.32µV)
(MOD: 1kHz/±1.5kHz)
2) Channel: 2
3) Channel: 3
4. Squelch
Open
(Wide)
1) Adj item: [SQOW]
Adjust:
[
✳✳✳
]
2) Adj item:
[L]
→
[C]
→
[H]
SSG
output:
12dB
SINAD level for
analog Wide –3.0dB
(MOD: 1kHz/±3kHz)
After input signal
from SSG, press
[PF5] key to store
the adjustment
value.
1) Adj item: [Squelch
Open (Wide)]
2) Adj item: [Low]
→
[Center]
→
[High]
SSG
output:
12dB
SINAD level for
analog Wide –3.0dB
(MOD: 1kHz/±3kHz)
After input signal
from SSG, press
[Apply] button to
store the adjust-
ment value.
[Panel
tuning
mode]
[PF5]
[PC test
mode]
[Apply]
[Panel tuning mode]
After input signal
from SSG, press
[PF5] key to store
the adjustment
value.
[PC test mode]
After input signal
from SSG, press
[Apply] key to store
the adjustment
value.
(Narrow)
1) Adj item: [SQON]
Adjust:
[
✳✳✳
]
2) Adj item:
[L]
→
[C]
→
[H]
SSG
output:
12dB
SINAD level for ana-
log Narrow –3.0dB
(MOD: 1kHz/±1.5kHz)
After input signal
from SSG, press
[PF5] key to store
the adjustment
value.
1) Adj item: [Squelch
Open (Narrow)]
2) Adj item: [Low]
→
[Center]
→
[High]
SSG
output:
12dB
SINAD level for ana-
log Narrow –3.0dB
(MOD: 1kHz/±1.5kHz)
After input signal
from SSG, press
[Apply] button to
store the adjust-
ment value.