Manual Operation
68
If “EXIT” is pressed, the display returns to the I/O-TEST menu.
4.5.10.2
CTRL OUT menu
In the submenu “CTRL OUT” a low or high level can be applied to the output.
I/O-TEST MENU
CTRL OUT
UP DOWN EXIT
P
SELECT OUTPUT
LEVEL !
HIGH LOW EXIT
Use the keys “HIGH” and “LOW” to set the output value or leave the menu with “EXIT”.
If “EXIT” is pressed, the display returns to the I/O-TEST menu.
4.5.10.3
ANALOG OUT menu
In the submenu “ANALOG OUT” a freely settable current between 2mA and 21mA can be applied to the output.
I/O-TEST MENU
ANALOG OUT
UP DOWN EXIT
P
ANALOG OUT TEST
OUTPUT 1
UP DOWN EXIT
Use the keys “UP” and “DOWN” to select the desired output and confirm with “P” or skip with “E”.
The display shows
ENTER CURRENT 1
00.0
mA
RIGHT UP EXIT
Use the keys “RIGHT” and “UP” to select the desired value and confirm with “P” or skip with “EXIT”.
If “P” was pressed the current is applied to the output and the display shows
ENTER NEW VALUE?
YES EXIT
Press “YES” to enter a new value or “EXIT” to leave the menu.
If “EXIT” is pressed, the display returns to the I/O-TEST menu.
4.5.10.4
CTRL IN menu
In the submenu “CTRL IN” the display shows the level currently applied to the control input.
I/O-TEST MENU
CTRL IN
UP DOWN EXIT
P
CURRENT CTRL-IN
LEVEL: LOW
. EXIT
After evaluating the input press “EXIT” to return to the I/O-TEST menu.
Summary of Contents for TRICOR TCE80 -E Series
Page 1: ...Certified according to DIN ISO 9001 Manual TRICOR Coriolis Mass Flow Meter ...
Page 2: ...SW Version Main V2 0x Display V2 0x Manual Revision 2 1 ...
Page 93: ...Listings 93 Fig 26 Dimensions TCE8000 E except A Fig 27 Dimensions TCE8000 E A ...
Page 98: ...Contact Copyright KEM Subject to change without notice ES Rev 002 23 01 2012 ...