Manual Operation
56
4.5.5 SETUP
Menu
In the SETUP menu all settings can be made to adapt the meter to the individual requirements.
Select in the main menu
MAIN MENU
SETUP
UP DOWN EXIT
Press “P”. The display shows
ENTER CODE
2206
LEFT UP EXIT
Change the indicated number with “LEFT” and “UP” to 2207 and confirm with “P”.
If a wrong code is entered, the display shows “ERROR” for about 2s and then asks for a new input.
When the correct code is entered the display shows:
SETUP MENU
PARAMETER
UP DOWN EXIT
The following submenus are available:
PARAMETER:
METER MODE
Selecting mass or volume meter mode
FLOW FILTER
Setting the filter time constant
CUT OFF
Setting the cutoff value
STEP RESP:
Adjusting the settings for fast changing flow
RESET KEY
Enable / disable the key “Reset”
FLOW-DIREC
Setting up the unit for reverse flow
K-FACTOR
Fine scaling the meter
FAULT TIME
Setting the error response time
LANGUAGE
Selecting the display language
IN/OUTPUTS:
FREQ OUT
Configuring the frequency output
CTRL OUT
Configuring the control output
ANALOG OUT
Configuring the analog output
CTRL IN
Configuring the control input
INTERFACE
Configuring the interface
DATA CONFIG:
SAVE DATA
Saving the actual settings as backup
RECALL DATA
recalling the last settings from the backup
RESET TOTAL:
Resets the total count to zero.
Use the keys “UP” and “DOWN” to select the desired submenu and confirm with “P” or skip with “E”.
Summary of Contents for TRICOR TCE80 -E Series
Page 1: ...Certified according to DIN ISO 9001 Manual TRICOR Coriolis Mass Flow Meter ...
Page 2: ...SW Version Main V2 0x Display V2 0x Manual Revision 2 1 ...
Page 93: ...Listings 93 Fig 26 Dimensions TCE8000 E except A Fig 27 Dimensions TCE8000 E A ...
Page 98: ...Contact Copyright KEM Subject to change without notice ES Rev 002 23 01 2012 ...