Manual Operation
65
CURRENTLY HART NOT
AVAILABLE, READ THE MANUAL
FOR MORE INFO
. OK
Confirm with “OK” and select an implemented interface.
4.5.8 SETUP DATA CONFIGURATION menu
In the submenu SETUP / DATA CONFIG the current setting can be stored to the backup memory and the stored
settings can be recalled
SETUP MENU
DATA CONFIG
UP DOWN EXIT
P
DATA CONFIG MENU
SAVE DATA
UP DOWN EXIT
Use the keys “UP” and “DOWN” to select the desired submenu and confirm with “P” or skip with “E”.
4.5.8.1 SAVE
DATA
menu
In the submenu “SAVE DATA” the current settings can be stored in the backup memory.
At each power on the TCE8000 compares the content of the setup memory and the backup memory. If the data in
those 2 memories are different, the TCE8000 gives out a warning. For avoiding that warning it is recommended to
make a backup as soon as the new settings are proven to be okay.
DATA CONFIG MENU
SAVE DATA
UP DOWN EXIT
P
READY TO SAVE DATA
START EXIT
Start the backup process with “START” or skip with “EXIT”.
If “START” is pressed, the display shows for some seconds
MEMORY
ACCESS
After that for about 2 seconds
READY
The display returns to the DATA CONFIG menu.