1-2
Performance Verification
Introduction
Use the procedures in this section to verify that Model 2420 accuracy is within the limits stated
in the instrument’s one-year accuracy specifications. You can perform these verification procedures:
•
When you first receive the instrument to make sure that it was not damaged during shipment.
•
To verify that the unit meets factory specifications.
•
To determine if calibration is required.
•
Following calibration to make sure it was performed properly.
WARNING
The information in this section is intended for qualified service personnel
only. Do not attempt these procedures unless you are qualified to do so.
Some of these procedures may expose you to hazardous voltages, which
could cause personal injury or death if contacted. Use standard safety pre-
cautions when working with hazardous voltages.
NOTE
If the instrument is still under warranty and its performance is outside specified limits, con-
tact your Keithley representative or the factory to determine the correct course of action.
Verification test requirements
Be sure that you perform the verification tests:
•
Under the proper environmental conditions.
•
After the specified warm-up period.
•
Using the correct line voltage.
•
Using the proper test equipment.
•
Using the specified output signals and reading limits.
Environmental conditions
Conduct your performance verification procedures in a test environment with:
•
An ambient temperature of 18-28°C (65-82°F).
•
A relative humidity of less than 70% unless otherwise noted.
Warm-up period
Allow the Model 2420 to warm up for at least one hour before conducting the verification
procedures.
If the instrument has been subjected to temperature extremes (those outside the ranges stated
above), allow additional time for the instrument’s internal temperature to stabilize. Typically, al-
low one extra hour to stabilize a unit that is 10°C (18°F) outside the specified temperature range.
Also, allow the test equipment to warm up for the minimum time specified by the manufacturer.
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...