![Keithley SourceMeter 3A 2420 Service Manual Download Page 54](http://html1.mh-extra.com/html/keithley/sourcemeter-3a-2420/sourcemeter-3a-2420_service-manual_3950476054.webp)
3-2
Routine Maintenance
Introduction
The information in this section deals with routine type maintenance that can be performed by
the operator.
Line fuse replacement
WARNING
Disconnect the line cord at the rear panel, and remove all test leads connect-
ed to the instrument (front and rear) before replacing the line fuse.
The power line fuse is accessible from the rear panel, just above the AC power receptacle (see
Figure 3-1).
Perform the following steps to replace the line fuse:
1.
Carefully grasp and squeeze together the locking tabs that secure the fuse carrier to the
fuse holder.
2.
Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1.
CAUTION
To prevent instrument damage, use only the fuse type specified in Table 3-1.
3.
Re-install the fuse carrier.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
250V
PEAK
75V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
90-250VAC
50, 60HZ
220VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
75V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V,
Ω
,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
3.15A, 250V
INTERLOCK-
DIGITAL I/O
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
Figure 3-1
Rear panel
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...