![Keithley SourceMeter 3A 2420 Service Manual Download Page 118](http://html1.mh-extra.com/html/keithley/sourcemeter-3a-2420/sourcemeter-3a-2420_service-manual_3950476118.webp)
B-6
Command Reference
Table B-2
Recommended :CALibration:PROTected:SENSe parameter ranges
Note: Actual allowable parameter ranges are ±50% of full range for zero parameters, 75% of full
scale to 150% of full scale for full-range parameters, but ranges listed above should be observed
for optimum calibration.
:SOURCE
(:CALibration:PROTected:SOURce)
Purpose
To calibrate the source function.
Format
:cal:prot:sour <nrf>
Parameters
See Table B-3 for recommended parameter ranges.
Description
The :CAL:PROT:SOUR command calibrates the Model 2420 source func-
tion. To use this command, you must (1) program the source to the correct
value, (2) select the range being calibrated, and (3) send the
:CAL:PROT:SOUR command for each parameter listed in Table B-3.
Note
When the Model 2420 receives this command, it will attempt to place the pa-
rameter into one of the four parameter ranges summarized in Table B-3, de-
pending on the active :SOUR:FUNC and range. If the parameter does not fit
into any of the four allowed ranges, an error number -222, "Parameter data
out of range" will be generated. Once the unit has successfully selected the
appropriate parameter range, it will then check to see if if the active source is
programmed to a value within the selected parameter range, and that
:OUTP:STAT is ON. If the active source is not programmed to a value within the
parameter range (or if the source is in autorange or in standby), an error num-
ber -221, "Settings conflict" will be generated. If no error occurs, the active
source function/range point will be calibrated using the corresponding pa-
rameter.
Sense
range
First parameter
(zero)
Second parameter
(negative full scale)
Third parameter
(positive full scale)
0.2V
2V
20V
60V
0.002 to +0.002
0.02 to +0.02
0.2 to +0.2
0.5 to +0.5
-0.18 to -0.22
-1.8 to -2.2
-18 to -22
-50 to -65
+0.18 to +0.22
+1.8 to +2.2
+18 to +22
+50 to +65
10
µ
A
100
µ
A
1mA
10mA
100mA
1A
3A
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
-3E-2 to 3E-2
-9E-6 to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3 to -11E-3
-90E-3 to -110E-3
-0.9 to -1.1
-2.7 to -3.15
+9E-6 to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3 to +11E-3
+90E-3 to +110E-3
+0.9 to +1.1
+2.7 to +3.15
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...