2-16
Calibration
Table 2-6
Recommended :CALibration:PROTected:SENse parameter ranges
Note: Parameter steps for each range may be performed in any order, but all three parameter steps for each
range must be completed. For optimum calibration, use parameters within recommended limits.
Table 2-7
Recommended :CALibration:PROTected:SOURce parameter ranges
Note: Parameter steps for each range may be performed in any order, but all four parameter steps for each range
must be completed. For optimum calibration, use parameters within recommended limits.
Sense
range
First parameter
(zero)
Second parameter
(negative full scale)
Third parameter
(positive full scale)
0.2V
-0.002 to +0.002
-0.18 to -0.22
+0.18 to +0.22
2V
-0.02 to +0.02
-1.8 to -2.2
+1.8 to +2.2
20V
-0.2 to +0.2
-18 to -22
+18 to +22
60V
-0.5 to +0.5
-50 to -65
+50 to +65
10
µ
A
-1E-7 to +1E-7
-9E-6 to -11E-6
+9E-6 to +11E-6
100
µ
A
-1E-6 to +1E-6
-90E-6 to -110E-6
+90E-6 to +110E-6
1mA
-1E-5 to +1E-5
-0.9E-3 to -1.1E-3
+0.9E-3 to +1.1E-3
10mA
-1E-4 to +1E-4
-9E-3 to -11E-3
+9E-3 to +11E-3
100mA
-1E-3 to +1E-3
-90E-3 to -110E-3
+90E-3 to +110E-3
1A
-1E-2 to +1E-2
-0.9 to -1.1
+0.9 to +1.1
3A
-3E-2 to +3E-2
-2.7 to -3.15
+2.7 to +3.15
Source range
Second parameter
(negative full scale)
Second parameter
(negative zero)
Third parameter
(positive full scale)
Fourth
parameter
(positive zero)
0.2V
-0.18 to -0.22
-0.002 to +0.002
+0.18 to +0.22
-0.002 to +0.002
2V
-1.8 to -2.2
-0.02 to +0.02
+1.8 to +2.2
-0.02 to +0.02
20V
-18 to -22
-0.2 to +0.2
+18 to +22
-0.2 to +0.2
60V
-50 to -65
-0.5 to +0.5
+50 to +65
-0.5 to +0.5
10
µ
A
-9E-6 to -11E-6
-1E-7 to +1E-7
+9E-6 to +11E-6
-1E-7 to +1E-7
100
µ
A
-90E-6 to -110E-6
-1E-6 to +1E-6
+90E-6 to +110E-6
-1E-6 to +1E-6
1mA
-0.9E-3 to -1.1E-3
-1E-5 to +1E-5
+0.9E-3 to +1.1E-3
-1E-5 to +1E-5
10mA
-9E-3 to -1E-3
-1E-4 to +1E-4
+9E-3 to +11E-3
-1E-4 to +1E-4
100mA
-90E-3 to -110E-3
-1E-3 to +1E-3
+90E-3 to +110E-3
-1E-3 to +1E-3
1A
-0.9 to -1.1
-1E-2 to +1E-2
+0.9 to +1.1
-1E-2 to +1E-2
3A
-2.7 to -3.15
-3E-2 to +3E-2
+2.7 to +3.15
-3E-2 to +3E-2
Summary of Contents for SourceMeter 3A 2420
Page 2: ...Model 24203A SourceMeter Service Manual A G R E A T E R M E A S U R E O F C O N F I D E N C E...
Page 15: ...1 Performance Verification...
Page 31: ...2 Calibration...
Page 53: ...3 Routine Maintenance...
Page 56: ...3 4 Routine Maintenance...
Page 57: ...4 Troubleshooting...
Page 70: ...4 14 Troubleshooting...
Page 71: ...5 Disassembly...
Page 78: ...5 8 Disassembly...
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ...6 Replaceable Parts...
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ...A Specifications...
Page 112: ...A 6 Specifications...
Page 113: ...B Command Reference...
Page 124: ...B 12 Command Reference...
Page 125: ...C Calibration Programs...