Disassembly 5-7
Instrument re-assembly
Re-assemble the instrument by reversing the previous disassembly procedures. Make sure
that all parts are properly seated and secured, and that all connections are properly made. To en-
sure proper operation, replace the analog signal wire ferrite noise filters, and securely fasten the
shield.
WARNING
To ensure continued protection against electrical shock, verify that power
line ground (green and yellow wire attached to the power module) is con-
nected to the chassis.
Also make sure the two bottom case screws are properly installed to secure
and ground the case cover to the chassis.