12-86 Instrument Control Library
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
smuX.source.compliance
X = SMU channel (a or b)
Attribute
Source compliance state.
Usage
compliance = smuX.source.compliance
Remarks
• Use this attribute to read the state of source compliance. true indicates that the limit
function is in control of the source (source in compliance). false indicates that the
source function is in control of the output (source not in compliance).
• This is a read-only attribute. Writing to this attribute will generate an error.
• Reading this attribute also updates the status model and the front panel with
generated compliance information.
Details
(basic operation) and
Also see
Example
Reads the source compliance state for SMU A:
compliance = smua.source.compliance
print(compliance)
Output: true
The above output indicates that the voltage limit has been reached (if configured as a
current source), or that the current limit has been reached (if configured as a voltage
source).