3-4
DUT Test Connections
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
Figure 3-2
Input/Output LO and chassis ground
Chassis
Screw
In/Out LO
(Chan A)
In/Out LOs are not connected
together and are electrically
isolated from chassis ground.
CHANNEL A
LO
G
LO
HI G G G HI
S
S
CHANNEL B
LO
G
LO
HI
G
G
G
HI
S
S
KEITHLEY Model 2602/2612 (rear panel)
Low-Noise
Chassis Ground
Banana Jack
In/Out LO
(Chan B)
terminals
Figure 3-3
FVR
Chassis
Series 2600
Signal
Ground
Chassis
Signal Ground is a local signal ground and defined as
the Low-Noise Chassis Ground Banana Jack.
Chassis is defined as the metal chassis of the Series 2600.
1) Frequency Variable Resistor (FVR) Isolates
the SMUs from high frequencies on the chassis.
For DC to 60Hz, the FVR is a virtual short (zero
ohms).
2)
DO NOT
use the Chassis Screw terminal to
make signal connections to external circuitry.
High frequency (>1MHz) on the chassis may
result in higher noise at the output.
Chassis Screw terminal is connected to the metal chassis of the
Series 2600.
1
Chassis
Screw
2
Low-Noise
Chassis Ground
Banana Jack
Signal
Ground
Low-Noise Chassis Ground Banana Jack and Chassis Screw