Series 2600 System SourceMeters Reference Manual
Source-Measure Concepts 8-11
Return to
2600S-901-01 Rev. A / May 2006
Figure 8-3
Model 2601/2602 continuous power operating boundaries
-V
+V
-40V
(IV)
Sink
+I
3A
1A
-6V
6V
40V
(III)
Source
(II)
Sink
-1A
-3A
-I
(I)
Source
-2.2A
-600mA
600mA
2.2A
Energy storage DUT
Energy storage DUT
£
30°C
T
amb
Model 2611/2612 continuous power operating boundaries
The general operating boundaries for Model 2611/2612 continuous power output
are shown in
(see “
General SourceMeter power equation
” in this sec-
tion for derating factors). In this drawing, the 1.5A, 20V and 100mA, 200V
magnitudes are nominal values. Also note that the boundaries are not drawn to
scale.