2-10 TSP Programming Fundamentals
Series 2600 System SourceMeters Reference Manual
2600S-901-01 Rev. A / May 2006
Return to
to be run. A user script can be saved in the non-volatile memory of the Series
2600. If it is not saved, the Series 2600 will lose the script when it is turned off.
Script management includes commands for the following operations:
• Retrieve scripts from non-volatile memory so they can be modified.
• Delete user scripts from non-volatile memory.
• Restore scripts in the run-time environment from non-volatile memory.
Figure 2-2
Run-Time
Environment
Non-Volatile
Memory
Loads
Scripts/Functions
On Power-Up
Create/Modify and Load Script
Save Any User Script
Returned Data
Run Any Script
Users Program
OR
Test Script
Builder
Script
Management
Programming
Options
KEITHLEY Series 2600
Host PC
Programming model for scripts