Performance Veri
fi
cation
Check DC Voltage
Accuracy with Remote
Sense
Equipment required
(Item 2) AC Power Source
(Item 5) High current hook up wire (bold line
connections)
(Item 3) Electronic Load
(Item 6) Low current hook up wire (
fi
ne line
connections)
(Item 4) DC Voltmeter
(Item 9) 0.050
Ω
, 5 W Resistor (Qty 2)
For each of the channels sequentially.
1.
Power off the DUT.
2.
Remove the shorting clip from the Remote Sense connector on the rear panel
for the channel being tested. Be sure to retain the clip or clips for the channels
not being tested.
Figure 1: Con
fi
guring the shorting clips on a Remote Sense connector to test
channel 1 on a 2230-30-1
Figure 2: Con
fi
guring the shorting clips on a Remote Sense connector to test
channel 2 on a 2230-30-1
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
27