Performance Veri
fi
cation
The performance veri
fi
cation procedures verify the performance of your
instrument. They do not adjust your instrument. If your instrument fails any of
the performance veri
fi
cation tests, you should contact Keithley service.
WARNING.
Observe all safety precautions listed in this manual before using
this product and any associated instrumentation. Although some instruments
and accessories are used with nonhazardous voltages, there are situations
where hazardous conditions may be present. This product is intended for use by
quali
fi
ed personnel who recognize shock hazards and are familiar with the safety
precautions required to avoid possible injury. Before using the product, carefully
read and follow all installation, operation, and maintenance information. Refer
to this manual for complete product speci
fi
cations. Before performing any
maintenance, disconnect the line cord and all test cables. Operators of this
instrument must be protected from electric shock at all times. The responsible
body must make sure that operators are prevented access and/or insulated from
every connection point. In some cases, connections must be exposed to potential
human contact. Product operators in these circumstances must be trained to
protect themselves from the risk of electric shock. If the circuit can operate at or
above 1000 volts, no conductive part of the circuit may be exposed.
WARNING.
Do not loosen any screw on this product. There are no user
serviceable components inside.
CAUTION.
Use properly rated load wires. All load wires must be heavy enough not
to overheat when carrying the maximum short-circuit output current of the power
supply. If there is more than one load, then any pair of load wires must be capable
of safely carrying the full-rated short-circuit output current of the power supply.
14
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference