Performance Veri
fi
cation
7.
Enter the voltmeter reading into the table for checking DC voltage accuracy
without remote sense. (See Table 16 on page 17.)
8.
Enter the voltmeter reading into the table for checking DC voltage readback
accuracy. (See Table 17 on page 18.)
9.
Enter the CUT Readback Voltage into the table for DC voltage readback
accuracy. (See Table 17 on page 18.)
10.
Calculate the difference of the two measurements taken in steps 8 and 9 and
enter the absolute value into the difference column of the table. (See Table 17
on page 18.)
11.
Increase the CUT output voltage by 25% of the FS output voltage.
12.
Repeat steps 7 through 11 until you complete testing at 100% of FS.
This completes the check for one channel. If needed, return to step 1 to run
through the check for the next channel.
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference
33